0703-001-700 Rev-C
EN
www.stryker.com
43
8.4.3.1 To Prepare the Docker
1. Make sure the docker power switch is in the ON
position and illuminated (Figure 44).
2. Make sure the bottle of Neptune Docking Detergent
REF 0700-001-026 is connected and has detergent to
perform a wash cycle.
NOTE:
The end of the inlet tube must be located at the
bottom of the detergent bottle. See the
instructions for use
supplied with the detergent bottle.
Figure 44 – To Prepare the Docker
8.4.3.2 To Perform a Docking Cycle
NOTE:
The default cycle will start within 15 seconds,
automatically, unless another option is manually selected
from the DOCKING MODE screen menu. See
Table 18
Docking Cycle Options
.
1. Push the rover toward the docker and between the
docker guides until the rover and docker attach
automatically (Figure 45). DO NOT lock the casters.
The DOCKING MODE screen will appear (Figure 15).
2. If another cycle option is selected before the initiation
of the default cycle, touch the START button to initiate
the selected cycle (Figure 46).
NOTE:
During the cycle a ‘progress’ message will appear
on the screen, followed by a ‘completed’ message.
3. After the cycle is complete, touch the RELEASE FROM
DOCK button to release the rover from the docker. Pull
the rover away from the docker.
4. Visually inspect the canisters for any remaining soil. If
soil remains, an Extended Wash cycle is available to
provide a more thorough cleaning of the canisters.
Figure 45 – To Dock the Rover
0000160552, Rev. C Effective Date: Nov 25, 2015 1:43:15 PM
Print Date: Nov 25, 2015 02:19:13 PM