13 Troubleshooting
13.1 Rover Operation
NOTE:
For service, contact your Stryker sales representative or call Stryker Neptune Customer Service. See
Section 1.3
Contact Information
. Outside the US, contact your nearest Stryker subsidiary.
SYMPTOM
PROBABLE CAUSE
ACTION
The rover does not power
up and the power switch is
in the ON position.
Power cord is not
connected or is not
connected securely.
Connect the supplied power cord or make sure the power cord is
connected securely.
Facility power is not
available or off.
Make sure facility power is provided to the hospital-grade power
receptacle.
The rover does not
recognize a manifold.
Manifold is altered or
damaged.
Replace the manifold. See
Section 5.3 For Use With
and the
instructions for use
supplied with the manifold.
No vacuum pump action
after the START SUCTION
button is touched.
Canisters are full and an
error occurs.
Dock the rover.
The rover is damaged.
Remove the rover from use. Contact Stryker Neptune Customer
Service. See
Section 1.3 Contact Information
.
The rover’s fluid suction is
weak or insufficient.
Manifold is not installed
correctly.
Rotate the manifold to make sure it is locked in place.
Unused manifold ports are
open.
Close all unused inlet ports of the manifold installed in each
canister.
Suction tubing connection
is not secure.
Make sure all suction tubing connections are secure.
Unused suction tubing is
not clamped.
Clamp any suction tubing not in use.
Suction tubing is blocked
or damaged.
Clear or replace any suction tubing in use.
Suction accessory is
blocked or damaged.
Clear or replace the suction accessory.
Suction tubing is too long
or has a narrow diameter.
Use shorter length or larger diameter suction tubing.
Manifold is clogged or
damaged.
Replace the manifold. See
Section 5.3 For Use With
and the
instructions for use
supplied with the manifold.
The suction control dials
are adjusted too low.
Turn the suction control dials to adjust the suction limit to the
proper level.
The fluid suction filter
requires replacement.
Replace the fluid suction filter. See
Section 5.3 For Use With
and
the
instructions for use
supplied with the filter.
The rover is damaged.
Remove the rover from use. Contact Stryker Neptune Customer
Service. See
Section 1.3 Contact Information
.
The rover’s suction
capability is lost.
An medium priority alarm
has occurred.
See
Section 13.3 Troubleshooting Codes.
The rover is damaged.
Remove the rover from use. Contact Stryker Neptune Customer
Service. See
Section 1.3 Contact Information
.