0703-001-700 Rev-C
EN
www.stryker.com
53
CODE MESSAGE
PRIORITY ACTION
12.X
Level
Sensor
Error
MEDIUM
Call service.
13.X
Vacuum
Error - 20L
MEDIUM
Call service.
13.6
20L
Canister
Pressure
LOW
Turn on suction to
reduce the positive
pressure in the
canister. If this
notification persists,
call service.
14.X
Vacuum
Error - 20L
MEDIUM
Call service.
15.X
Vacuum
Error - 4L
MEDIUM
Call service.
15.6
4L Canister
Pressure
LOW
Turn on suction to
reduce the positive
pressure in the
canister. If this
notification persists,
call service.
16.X
Vacuum
Error - 4L
MEDIUM
Call service.
18.3
4L Almost
Full
LOW
Prepare to use an
alternate source of
suction, or empty the
4L canister of fluid
waste to prevent the
loss of suction.
18.4
20L Almost
Full
LOW
Prepare to use an
alternate source of
suction. 20L suction
will stop when the 20L
canister is full.
18.5
4L Full
MEDIUM
Use an alternate
source of suction, or
empty the 4L canister
to restore lost suction
to the 4L canister.
18.6
20L Full
MEDIUM
Use an alternate
source of suction,
or dock the rover
to empty the 20L
canister.
CODE MESSAGE
PRIORITY ACTION
20.1
Audio
Failure
MEDIUM
Cycle power to the
rover. Push the power
switch to the OFF
position; then push
the switch to the ON
position. If this error
persists, call service.
21.0
Knob/Dial
Error
MEDIUM
Call service.
22.0
4L
Manifold
Error
MEDIUM
Call service.
22.1
22.2
23.0
20L
Manifold
Error
MEDIUM
Call service.
23.1
23.2
24.11 4L
Manifold
Error
LOW
Install new manifold.
24.13
Expired 4L
Manifold
LOW
Install new manifold.
24.15
Used 4L
Manifold
LOW
Install new 4L manifold
before performing
the next surgical
procedure.
25.11 20L
Manifold
Error
LOW
Install new manifold.
25.13
Expired
20L
Manifold
LOW
Install new manifold.
25.15
Used 20L
Manifold
LOW
Install new 20L
manifold before
performing the next
surgical procedure.
0000160552, Rev. C Effective Date: Nov 25, 2015 1:43:15 PM
Print Date: Nov 25, 2015 02:19:13 PM