14 Specifications
Model:
Neptune 3 Rover
REF:
0703-001-000
Electrical Power
Requirements:
120 V
, 60 Hertz (Hz), 12.0 Amps (A), single phase
20 V
, 3 A during docking procedure; rover receives power from docker REF 0702-014-000
Product Safety
Certification:
Canadian Standards Association (CSA) International
American National Standards Institute (ANSI)/Association for the Advancement of Medical
Instrumentation (AAMI)
ANSI/AAMI ES60601-1:2005/(R) 2012 and A1: 2012, Medical Electrical Equipment — Part 1:
General Requirements for Basic Safety and Essential Performance
; Consolidated Reprint
(2009); Amendment 2 (2010); 3rd Edition, 3.1, includes Amendment 1 (2012).
Product Safety
Compliance:
International Electrotechnical Commission (IEC)
IEC 60601-1:2005, Medical Electrical Equipment — Part 1: General Requirements for Basic
Safety and Essential Performance
; IEC Corrigendum 1 (2006); IEC Corrigendum 2 (2007) +
AM1:2012
IEC 60601-1-6:2010 (Third Edition) + A1:2013, Medical Electrical Equipment – Part 1-6: General
Requirements for Basic Safety and Essential Performance
IEC 62366-1:2007 (First Edition) + A1:2014, Medical Devices - Part 1: Application of Usability
Engineering to Medical Devices
IEC 60601-1-8:2006 (Second Edition) + AM1:2012, Medical Electrical Equipment – Part 1-8:
General Requirements for Basic Safety and Essential Performance – Collateral Standard:
General Requirements, Tests and Guidance for Alarm Systems in Medical Electrical Equipment
and Medical Electrical Systems
IEC 62304:2006 (First Edition), Medical Device Software – Software Life Cycle Processes
IEC 62471:2006 (First Edition), Photobiological safety of lamps and lamp systems
Dimensions:
Width:
48.3 cm [19 inch]
Height:
259 cm [102 inch] with powered IV pole at maximum height; 177.8 cm [70 inch] with powered
IV pole at minimum height
Depth:
58.4 cm [23 inch]
Mode of Operation:
Continuous
Sound Pressure:
Medium Priority Alarm, 57 - 62 dB (not adjustable, see
IEC 60601-1-8: 2006 (Second Edition) +
AM1:2012
)
Adjustable Suction Limit:
50 to 520 mm-Hg; measured with all ports closed
Vacuum Measurement
Accuracy:
± 5% of full scale (± 26 mm-Hg)
Suction Limit Accuracy:
< 26 mm-Hg or 10% of setting