0703-001-700 Rev-C
EN
www.stryker.com
45
9 Inspection and Maintenance
WARNINGS:
▪ Upon initial receipt and before each use, inspect each
component for damage. DO NOT use any equipment if
damage is apparent. See
Table 19 Inspection Schedule
and Criteria.
▪ DO NOT disassemble, modify, service, or repair
any equipment without the authorization of the
manufacturer.
NOTES:
▪ Only biomedical equipment technicians trained and
experienced in the maintenance of this reusable
medical device should install and maintain this
equipment.
▪ For service, contact your Stryker sales representative
or call Stryker Neptune Customer Service. See
Section
1.3 Contact Information
. Outside the US, contact your
nearest Stryker subsidiary.
▪ Maintenance documentation for this equipment is
available upon request to Stryker-authorized service
personnel only.
Table 19 – Inspection Schedule and Criteria
INTERVAL
CRITERIA
ACTION
Before initial use
Inspect
equipment
for damage
or missing
components.
If damage is
apparent, replace
the equipment.
See
Section 5.3
For Use With
.
Make sure the
rover and docker
operate properly.
See the
following:
Section
8.4.3 To Dock the
Rover, Section
8.1.3 To Test
the Rover, and
Section 8.1.4 To
Adjust the Rover
Settings.
Before each use
and after each
cleaning and
disinfection
Inspect
equipment
for damage
or missing
components.
If damage is
apparent, replace
the equipment.
See
Section 5.3
For Use With
.
INTERVAL
CRITERIA
ACTION
Before each use
and after each
cleaning and
disinfection
Inspect for
corrosion,
discoloration,
pitting, cracked
materials, or
unacceptable
deterioration
on any external
surfaces,
including labels
and product
markings.
If damage is
apparent, replace
the equipment.
See
Section 5.3
For Use With
.
Inspect the
canister exterior,
smoke evacuator
filter cover,
and infrared
communication
window for
cracks or
damage.
Inspect all of
the casters and
make sure the
locks function
properly.
Inspect the
power cord for
cuts and the
power cord plug
for bent pins.
Inspect the
power cord
receptacle for
bent pins or bent
contacts.
Inspect the
infrared
communication
window for any
dirt or debris.
Thoroughly
remove any
dirt or debris
on the infrared
communication
window. See
Section 10
Cleaning and
Disinfection
.
0000160552, Rev. C Effective Date: Nov 25, 2015 1:43:15 PM