40
Error message
Possible cause
Remedy
No function
Power plug not connected to outlet
Check power plug and, if necessary, connect to power supply.
Emergency Laser Emission Stop Switch activated
Unlock emergeny stop switch by turning it.
Red LED “Fiber” activated
Fiber disconnected
Connect fiber or inspect it.
E 03
Footswitch detective
Check footswitch and connect it, if necessary.
E 04
Release footswitch
Disconnect and then reconnect footswitch and switch on laser again; if
necessary, replace/send in footswitch for repair.
E 05
Door interlock
Reconnect remote interlock.
E 06
Flash lamp aged
Note error message and send in unit for servicing.
Replace flash lamp.
E 07
Flash lamp detective
Send in unit for servicing and replace flash lamp. You may not operate the
laser again until the flash lamp has been replaced.
E 10
Bad quality of cooling water
Arrange date for servicing and change cooling water.
E 11
Laser on time out of specifications
Let unit run in standby mode for 5 mins.
E 14
Cooling water temperature too low.
Wait until heating-up phase is complete.
E 15
Cooling temperature to high
Let unit cool in standby mode, check room temperature and lower it, if
necessary.
E 22
Simmer default
Note error code; unit normally reignites simmer within approx. 5 sec. and
goes into standby; if operation fails (unit switches itself off automatically)
arrange servicing date.
E 23 - 40
Note error code and arrange servicing date.
Technische Beschreibung
Fehlersuchliste
Technical Description
Troubleshooting
Техническое описание
Выявление неисправностей
Содержание 27750120-1
Страница 1: ...CALCULASE CALCULASE CALCULASE 27750120 1 27750120 1 27750120 1...
Страница 2: ......
Страница 3: ...Gebrauchsanweisung Instruction Manual CALCULASE CALCULASE CALCULASE 27750120 1 27750120 1 27750120 1...
Страница 5: ...III...
Страница 6: ...IV W E Q R U Y T q w e P O I r t y p i u o neue Abbildung...
Страница 10: ......
Страница 55: ...45 Technische Beschreibung Technical description Blockschaltbild Block diagram...
Страница 75: ......