35
Table 202
Guidance and manufacturer’s declaration – electromagnetic immunity
HALOGEN 250-2 twin model
20
1134 20 is intended for use in the electromagnetic environment specified below. The customer or user of the
HALOGEN 250-2 twin model
20
1134 20 should ensure that it is used in such an environment.
Immunity test
EN IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
Complies
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
Complies
± 2 kV for power supply
lines
± 1 kV for input/output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
Complies
± 1 kV differential mode
± 2 kV common mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% U
T
*
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
Complies
<5% U
T
*
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 0.5 cycle
See guidelines
70% U
T
(30% dip in U
T
)
for 1 cycle
See guidelines
See guidelines
Mains power quality should be that of a typical
commercial or hospital environment.
If the user of the equipment requires continued
operation during voltage dips, short interruptions,
voltage variations on power supply input lines, or
operation at <5% U
T
(>95% dip in U
T
) for 5 seconds, it
is recommended that the equipment or system be
powered from an uninterruptible power supply.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
Complies
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
* Note:
U
T
is the a.c. mains voltage prior to application of the test level.