10 Alarms and troubleshooting
48
10.5 Sensor defect alarm A5
(medium priority alarm)
Control panel
Alarm condition
At least one temperature sensor supplies a value outside
the permissible range.
Device response
A5
(flashing) appears on the display (5). If both outputs
are active,
A5
is displayed for both outputs.
The Start LED (6) and the Alarm LED (10) flash.
An acoustic alarm signal is triggered.
The heating process is interrupted at the two outputs.
Possible reasons
►Required
action(s)
Defective sensor(s) in the applied part.
►Send the applied part to the local sales point.
Defective connection cable on the applied part.
►Send the applied part to the local sales point.
Defective cable plug on the applied part.
►Send the applied part to the local sales point.
Defective extension connection cable.
►Replace the extension connection cable.
Defective connecting socket for output A or B on the con-
trol unit.
►Send the control unit to the local sales point.