10 Alarms and troubleshooting
ASTOPAD
®
Instructions for Use
49
10.6 Heating defect alarm A6
(medium priority alarm)
Control panel
Alarm condition
The heater of the applied part is defect.
Device response
A6
(flashing) appears on the display (5). If both outputs
are active,
A6
is displayed for both outputs.
The Start LED (6) and the Alarm LED (10) flash.
An acoustic alarm signal is triggered.
The heating process is interrupted at the two outputs.
Possible reasons
►Required
action(s)
Heater resistance too high or low.
►Send the applied part to the local sales point.
Defective connection cable on the applied part.
►Send the applied part to the local sales point.
Defective cable plug on the applied part.
►Send the applied part to the local sales point.
Defective extension connection cable.
►Replace the extension connection cable.
Defective connecting socket for output A or B on the con-
trol unit.
►Send the control unit to the local sales point.