ASTOPAD
®
Instructions for Use
13 Maintenance
56
The inspection of the Essential Performance characteristics and all other safety-
relevant functions are described in the ASTOPAD testimg instructions:
Testing instructions ASTOPAD, German (Order no. 1731.9045.11)
Testing instructions ASTOPAD, English (Order no. 1731.9045.12)
The following test equipment is required to carry out the repeated inspection:
System test box for ASTOPAD (Order no. 1715.9040)
The test instructions are included in the scope of delivery of the ASTO-
PAD system test box.
13.2 Replacing the battery
1. Disconnect the ASTOPAD control unit completely from the power supply (by
pulling out the mains plug).
2. Press the Standby button until the Standby LED goes out.
3. Disconnect the mains connection cable from the control unit, flipping the lock
(red button) of the mains plug in the direction of pull.
4. Remove the housing cover (Fig. 6) by releasing the four housing screws.
5. Remove the battery (Fig. 7) and insert a new battery as far as it will go.
6. Replace the housing cover and secure with the four housing screws.
Fig. 6 Housing cover
Fig. 7 Battery
Battery