7 Operating states
ASTOPAD
®
Instructions for Use
25
7 Operating states
With a few exceptions, the operating states when only one applied part is con-
nected to the ASTOPAD control unit at output A and is being operated are de-
scribed below.
The description of the operating states applies to the connection and operation of
a second applied part to output B of the ASTOPAD control unit.
7.1 Standby mode
Control panel
Action
After connecting the mains plug to the mains socket, the con-
trol unit is in
Standby mode
.
Device re-
sponse
The Standby LED (9) lights up.
The display (11) shows the symbol for battery status (only
for equipment with built-in battery).
Press the Standby button (9) to switch the equipment from
any mode to
Standby mode
.
In
Standby mode
only the electronics and the applied parts
are disconnected from the power supply. The control unit
remains connected to the mains.
For equipment with a built-in battery, the battery is charged
in
Standby mode
.
After a power failure, the equipment automatically switches
to
Standby mode
.