14 Technical Data
ASTOPAD
®
Instructions for Use
59
ASTOPAD
SOF2
SOF4
SOF5
SOF7
Power consumption (W)
105
115
150
60
Dimensions* (mm)
Length
(with sacral)
Width
Height
approx.
1710-2300
480-600
40-100
approx.
810-1300
(910-1300)
480-600
40-100
approx.
1310-1700
480-600
40-100
approx.
600-800
(660-900)
450-600
40-100
Weight (kg)
Approx. 7.5 Approx. 2.7 Approx. 3.2 Approx. 2.0
Connection cable
50 cm PVC cable
Standard extension con-
nection cable
2 m
* according to ordering information
ASTOPAD
ROE4
ROE8
Power consumption (W)
105
Dimensions* (mm)
Length
Height
approx.
2400-2800
40
approx.
2400-2800
80
Weight (kg)
Approx. 6.5
Approx. 13.0
Connection cable
200 cm PVC cable
* according to ordering information
ASTOPAD
Integrable rechargeable battery for the
ASTOPAD DUO310 control unit (optional)
Type
Li-Ion
Energy content (Wh)
99.4
Dimensions (mm)
Length x width x height
150 x 77 x 22
Weight (g)
430
NOTICE
The specified defibrillation protection is ensured only when the applied part is con-
nected to the extension connection cable and the control unit.
DANGER
Explosion hazard!
Do not use the ASTOPAD patient warming system in an explosive environment or
in the presence of flammable anaesthetics.