17 Guidelines and manufacturer’s declaration
ASTOPAD
®
Instructions for Use
63
17 Guidelines and manufacturer’s declaration
Guidelines and manufacturer's declaration - electromagnetic emissions
ASTOPAD is intended for use in the electromagnetic environment specified below. The customer or user of ASTOPAD
should assure that the system is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidelines
RF emissions in accordance
with CISPR 11/EN 55011
Group 1
ASTOPAD uses RF energy only for its internal function.
Therefore, its RF emissions are very low, and it is unlikely that
nearby electronic devices will be affected.
RF emissions in accordance
with CISPR 11/EN 55011
Class A
The emission characteristics of this device permit its use in in-
dustrial and hospital environments (CISPR 11, Class A). If
used in domestic settings (which usually requires Class B ac-
cording to CISPR 11), this device may not offer suitable pro-
tection of radio services. The user may have to take corrective
action such as relocation or re-orientation of the equipment.
Harmonic components in ac-
cordance with IEC 61000-3-2
Class A
Voltage fluctuations and flicker
in accordance with
IEC/EN 61000-3-3
Complies
Guidelines and manufacturer's declaration - electromagnetic immunity
ASTOPAD is intended for use in the electromagnetic environment specified below. The customer or user of ASTOPAD
should assure that the system is used in such an environment.
Immunity test
Test level
Compliance level
Electromagnetic environment - guidelines
Electrostatic dis-
charge (ESD) ac-
cording to
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
Complies
The floor should be a wood, concrete, or ceramic
tile floor. If the floor is made of synthetic material,
the relative humidity must be at least 30%.
Electrical fast transi-
ents and bursts ac-
cording to
IEC 61000-4-4
± 2 kV
100 kHz repetition
frequency
Complies
Mains power quality should be that of a typical
commercial or hospital environment.
Surges according to
IEC 61000-4-5
± 0.5 kV, ± 1 kV
Line to line
± 0.5 kV, ± 1 kV,
± 2 kV line to
ground
Complies
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips in ac-
cordance with
IEC 61000-4-11
0 % U
T
; ½ cycle
At 0, 45, 90, 135,
180, 225, 270 and
315 degrees
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30
cycles
Single-phase at
0 degrees
Complies
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the equipment requires continued operation even
in the event of mains power interruptions, it is rec-
ommended that the equipment be powered from
an uninterruptible power supply or battery.
Voltage interruptions
in accordance with
IEC 61000-4-11
0 % U
T
; 250/300
cycles
Complies
Magnetic fields with
energy rated fre-
quencies according
to
IEC/EN 61000-4-8
30 A/m
50 Hz or 60 Hz
Complies
Power frequency magnetic fields should be at lev-
els characteristic of a typical commercial or hospi-
tal environment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.