ASTOPAD
®
Instructions for Use
9 Getting started
40
9.5 Switching off the ASTOPAD
1. Press the Standby
button of the ASTOPAD control unit to switch off (all
displays are turned off, the Standby LED
illuminates).
2. Disconnect the ASTOPAD control unit from the applied part(s).
Operate the ASTOPAD applied part ROE:
To achieve optimum heating conditions for the next patient, end the heat-
ing process after each use.
9.6 Cleaning and disinfection
During use, ASTOPAD and its accessories may be unintentionally contaminated
with organic contaminants (such as blood or bodily fluids) or micro-organisms.
The following cleaning and disinfection procedures should therefore be followed
after each use.
Wear gloves when cleaning and disinfecting (chemical-resistant).
Always work from top to bottom and from clean to dirty areas.
When using wipes, only use for as long as they leave a closed film of fluid on
the surface.
Please observe the defined contact time of the disinfectant.
Work methodically to disinfect each area of the ASTOPAD.
9.6.1 Preparation
For routine disinfection, only use alcohol-based disinfectants or one of the follow-
ing approved disinfectants:
Disinfectant
Manufacturer
acryl-des
Schülke + Mayr, Austria
ANIOSURF CITRON
Laboratoires ANIOS, France
Bacillol Plus
BODE CHEMIE HAMBURG, Germany
BIGUAMED PERFEKT N
Desomed-Dr. Trippen GmbH, Germany
ClearSurf
Fresenius Medical Care, Germany
Mikrobac forte
BODE CHEMIE HAMBURG, Germany
mirkrozid sensitive liquid
Schülke + Mayer, Germany
Terralin protect
Schülke + Mayer, Germany
Incidin Plus
Ecolab GmbH, Germany
Incidin Pro
Ecolab GmbH, Germany
Meliseptol Foam pure
B. Braun Melsungen AG, Germany