6-7
Sterilizer Operation
Operator Manual
P129394-020
6.6 Loading Car
Instructions:
Unloading
NOTE: The loading car and transfer carriage are for use with the 20-inch
AMSCO 400 Series sterilizer, only.
1.
Open sterilizer door.
2.
Move transfer carriage forward until latches engage with track
inside chamber.
3.
Verify that transfer carriage latched to track inside chamber by
pulling transfer carriage backward (transfer carriage should
remain stationary).
4.
Once transfer carriage is latched to track inside chamber, grasp
the loading car and carefully pull loading car from chamber onto
transfer carriage until transfer carriage latch engages to loading
car.
5.
Disengage transfer carriage latches from track inside chamber
by pushing carriage latch knob.
6.
Close the sterilizer door.
7.
Transfer load from sterilizer area.
6.7 Loading/Unloading
Sterilizer:
Rack and Shelves
If sterilizer is equipped with the rack and shelves option, refer to
instructions below and
Figure
1.
Open chamber door.
2.
Transfer load to shelves in chamber. Shelves slide out halfway to
facilitate loading.
3.
After loading the shelves, slide them to closed position to verify
shelf does not interfere with door operation (both doors if double-
door sterilizer).
4.
Close chamber door(s).
5.
The sterilizer is now ready to run a cycle. Refer to appropriate
cycle description in
section of the manual).
WARNING – BURN HAZARD:
Sterilizer, rack/shelves and
loading car will be hot after
cycle is run. Always wear
protective gloves and apron
when removing a processed
load. Protective gloves and
apron must be worn when
reloading sterilizer following
the previous operation.
!
WARNING – BURN HAZARD:
Steam may be released from
the chamber when door is
opened. Step back from the
sterilizer each time the door is
opened to minimize contact
with steam vapor.
WARNING – PERSONAL
INJURY HAZARD: When
closing the chamber door,
keep hands and arms out of
the door opening and make
sure opening is clear of
obstructions.
!
!
WARNING – BURN HAZARD:
•
Reduce risk of liquid boil-
over. Do not place liquid-
filled bottles within 4”
(102 mm) of chamber door
during processing.
•
Sterilizer, rack/shelves and
loading car will be hot after
cycle is run. Always wear
protective gloves and apron
when removing a processed
load. Protective gloves and
apron must be worn when
reloading sterilizer following
the previous operation.
!
Содержание AMSCO 400 Series
Страница 2: ......
Страница 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Страница 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Страница 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Страница 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 80: ...6 20 P129394 020 Operator Manual Sterilizer Operation Figure 6 10 Typical Printout of a Liquid Cycle ...
Страница 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...