9-17
Troubleshooting
Operator Manual
P129394-020
Alarm
Description
Screen with Operator Instructions
9.3.5 Too Long To
Open Door
Occurs if door switch
does not open within the
allotted time.
Screen with Service Instructions
STATUS . . ALARM!
245
TOO LONG TO OPEN DOOR
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
REMAIN IN ALARM CONDITION UNTIL
DOOR IS OPENED
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
246
TOO LONG TO OPEN DOOR
-> DOOR SWITCH DID NOT OPEN IN ALLOTTED TIME
CAUSES AND CORRECTIONS:
1. DOOR SWITCH MALFUNCTION
-> CHECK LS1 (LS2) CONNECTIONS
-> READJUST LS1 (LS2)
2. POWER DOOR MECHANISM FAILURE
-> REPAIR MECHANISM
-> REPLACE MOTOR
PAPER
FEED
EXIT
CLEAR
ALARM
Содержание AMSCO 400 Series
Страница 2: ......
Страница 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Страница 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Страница 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Страница 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 80: ...6 20 P129394 020 Operator Manual Sterilizer Operation Figure 6 10 Typical Printout of a Liquid Cycle ...
Страница 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...