8-3
Routine Maintenance
Operator Manual
P129394-020
8.3 Routine
Maintenance
Procedure
Full information for chamber cleaning can be found in STERIS
Technical Bulletin, M2826.
8.3.1 Flush Chamber Drain
Recommended Frequency:
Flush chamber drain whenever the line
becomes clogged:
1.
Turn off steam supply valve. Wait until jacket pressure is zero.
Wait until chamber has cooled to room temperature.
2.
Remove chamber drain strainer (
Figure
). Clean strainer
using procedures given
, if necessary.
3.
Pour a solution of 60 mL (1/4 cup) of Liqui-Jet 2 Instrument
Detergent (
Contact your local STERIS representative
), and 500 mL
(1 pint) of hot water into the drain. Solution may puddle in the
bottom of the chamber.
Should the detergents in
Step 3
be unavailable, you may use a
hot solution of 15mL (1 tablespoon) of trisodium phosphate to
500 mL (1 pint) of hot water.
4.
Open door and place strainer back in drain.
8.3.2 Chamber Cleaning
Recommended Frequency:
Clean chamber whenever it becomes
soiled, or as needed to maintain clean appearance.
NOTE: To clear any puddling of cleaning solution in the bottom of the
chamber, any cycle can be run, then aborted following the cycle’s
Purge phase.
For additional information, please refer to
T
ECHNICAL
B
ULLETIN
,
STERIS S
TEAM
S
TERILIZATION
C
LEANING
P
ROCEDURE
(M2826),
available directly from STERIS.
Due to differences in water quality, steam quality, frequency of use
and boiler additives, no specific frequency interval is recommended
for chamber cleaning. Operators must determine appropriate
cleaning interval based on the local water conditions and chamber
appearance.
Chamber cleaning should be performed:
•
When spills or other soiling have occurred in the chamber
•
Periodically to maintain sterilizer cleanliness and appearance
•
When processing materials that may be detrimental to the
sterilizer (e.g., chloride-containing substances)
1.
If applicable, the shelf assembly must be removed before
cleaning the chamber.
Single Door
a.
Remove shelves from rack.
b. Use an 1/8" hex wrench to loosen (but not remove) the set
screws at the front of the rack assembly.
c.
Remove the rack assembly from the chamber.
Double Door
a.
Remove shelves from rack.
b. Use an 1/8" hex wrench to loosen (but not remove) the set
screws at each end of the rack assembly.
c.
Remove the rack assembly from the chamber.
WARNING – BURN HAZARD:
Allow sterilizer and
accessories to cool to room
temperature before
performing any cleaning or
maintenance procedures.
!
WARNING – SLIPPING
HAZARD: To prevent falls,
keep floors dry by
immediately wiping up any
spilled liquids or
condensation in sterilizer
loading and unloading areas.
WARNING – BURN HAZARD:
Allow sterilizer to cool to room
temperature before
performing any cleaning or
maintenance procedures.
!
!
Содержание AMSCO 400 Series
Страница 2: ......
Страница 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Страница 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Страница 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Страница 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 80: ...6 20 P129394 020 Operator Manual Sterilizer Operation Figure 6 10 Typical Printout of a Liquid Cycle ...
Страница 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...