![Steris AMSCO 400 Series Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/steris/amsco-400-series/amsco-400-series_operators-manual_1374315059.webp)
5-23
Component Identification
Operator Manual
P129394-020
4.
Complete
– The automated generator drain ball valve closes at
end of flush.
•
If
Auto Utility
mode is enabled, the generator chamber fills
and begins heating water at time of day utilities are set to
restart.
•
If
Manual Utility
mode is enabled:
»
If flush begins while sterilizer is in Standby mode at
scheduled time, sterilizer returns to Standby when flush
is complete. Utilities restart when operator selects ON to
leave standby screen.
»
If flush began from main menu screen, sterilizer returns
to main cycle screen and utilities restart immediately.
NOTE: If sterilizer is in-cycle at scheduled flush time, a
flush cycle automatically occurs when sterilization cycle is
completed. Unload items from chamber then close door
to aid in depressurize phase.
•
Approximately 10 – 15 minutes are required to build a head
of steam suitable for running a sterilization cycle.
Important:
After removing load(s) from the chamber, close the
chamber door and keep the chamber door closed to minimize utility
consumption.
STATUS . .FLUSH 0:00
4
TEMP . . .00 F
PRESS . . .00 inHg
GENERATOR FLUSH
PROJECTED CYCLE COMPLETION TIME:
05:00
MINUTES
SECONDS
PAPER
STATUS
FEED
ABORT
Содержание AMSCO 400 Series
Страница 2: ......
Страница 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Страница 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Страница 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Страница 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 80: ...6 20 P129394 020 Operator Manual Sterilizer Operation Figure 6 10 Typical Printout of a Liquid Cycle ...
Страница 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...