9-16
P129394-020
Operator Manual
Troubleshooting
Alarm
Description
Screen with Service Instructions
Atmospheric Pressure
Alarm (Continued)
Alarm
Description
Screen with Operator Instructions
9.3.4 Too Long To
Close Door
Occurs if door switch
does not make contact
within allotted time.
Screen with Service Instructions
STATUS . . SERVICE INFORMATION:
270
ATMOSPHERIC PRESSURE ALARM
-> THE ATMOSPHERIC PRESSURE IS
GREATER THAN 1 PSI FROM THE
CALIBRATED ATMOSPHERIC PRESSURE
CAUSES AND CORRECTIONS:
1. THE STERILIZER IS AT A HIGHER
ALTITUDE THEN WHERE IT WAS
ORIGINALLY CALIBRATED (FACTORY)
-> SET ALTITUDE
2. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
3. LOOSE CONNECTION IN TRANSDUCER WIRING
-> REPAIR
4. TRANSDUCER FAILED
-> REPLACE
-> RECALIBRATE
PAPER
FEED
EXIT
CLEAR
ALARM
STATUS . . ALARM!
239
TOO LONG TO CLOSE DOOR
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
REMAIN IN ALARM CONDITION UNTIL
DOOR IS CLOSED
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK DOOR FOR OBSTRUCTION
-> REMOVE OBSTRUCTION AND CLOSE DOOR
3. IF DOOR WILL NOT CLOSE, CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
240
TOO LONG TO CLOSE DOOR
-> DOOR SWITCH DID NOT MAKE IN ALLOTTED TIME
CAUSES AND CORRECTIONS:
1. DOOR SWITCH MALFUNCTION
-> CHECK LS1 (LS2) CONNECTIONS
-> READJUST LS1 (LS2)
-> REPAIR LS1 (LS2)
2. POWER DOOR MECHANISM FAILURE
-> REPAIR MECHANISM
-> REPLACE MOTOR
PAPER
FEED
EXIT
CLEAR
ALARM
Содержание AMSCO 400 Series
Страница 2: ......
Страница 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Страница 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Страница 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Страница 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 80: ...6 20 P129394 020 Operator Manual Sterilizer Operation Figure 6 10 Typical Printout of a Liquid Cycle ...
Страница 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...