9-5
Troubleshooting
Operator Manual
P129394-020
Alarm
Description
Screen with Operator Instructions
9.2.4 Door Unsealed
Occurs if steam
pressure in door seal
drops below 5 psig.
Screen with Service Instructions
Alarm
Description
Screen with Operator Instructions
9.2.5 Chamber
Pressure/Temperature
Failure
Two conditions can
cause this type of alarm:
1. Chamber
temperature is
greater than 250°F
(121°C) and pressure
is less than 11 psig
or
Chamber
temperature is 270°F
(132°C) and pressure
is less than 20 psig.
2. Chamber and Jacket
temperatures are
more than 10°F apart.
STATUS . . ALARM!
207
DOOR UNSEALED
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
•
EXHAUST OR AIR BREAK CHAMBER TO
ATMOSPHERIC PRESSURE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
208
DOOR UNSEALED
-> STEAM PRESSURE IN DOOR SEAL BELOW
10 PSIG
CAUSES AND CORRECTIONS:
1. SEAL PRESSURE SWITCH MALFUNCTION
-> CHECK PS1 (PS2) CONNECTIONS
-> READJUST PS1 (PS2)
-> REPAIR PS1 (PS2)
2. SEAL NOT ACTIVATED
-> CHECK SEAL STEAM
-> CHECK SEAL EXHAUST
3. SOLENOID VALVE MALFUNCTION
-> REPAIR S35 (S36)
PAPER
FEED
EXIT
CLEAR
ALARM
STATUS . . ALARM!
219
PRESSURE/TEMPERATURE FAILURE
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
Содержание AMSCO 400 Series
Страница 2: ......
Страница 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Страница 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Страница 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Страница 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 80: ...6 20 P129394 020 Operator Manual Sterilizer Operation Figure 6 10 Typical Printout of a Liquid Cycle ...
Страница 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...