9-3
Troubleshooting
Operator Manual
P129394-020
Alarm
Description
Screen with Operator Instructions
9.2.2 Too Long In
Evacuation
Occurs if chamber does
not reach the set
evacuation level within
the allotted time.
Screen with
Service
Instructions
NOTE: This alarm has
two service help screens.
STATUS . . ALARM!
202
TOO LONG IN EVACUATION
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPETE CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK WATER SUPPLY VALVE
-> IF CLOSED, OPEN VALVE
3. IF ALARM RECURS, ABORT CYCLE
AND CALL SERVICE
SERVICE
HELP
ABORT
SILENCE
ALARM
STATUS . . SERVICE INFORMATION:
204
TOO LONG IN EVACUATION
-> CHAMBER DID NOT REACH REQUIRED
VACUUM LEVEL WITHIN ALLOTTED TIME
CAUSES AND CORRECTIONS:
1. LEAK IN PLUMBING
-> REPAIR
-> RUN A LEAK TEST
2. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
EXIT
PAPER
FEED
ABORT
STATUS . . SERVICE INFORMATION:
203
TOO LONG IN EVACUATION
-> CHAMBER DID NOT REACH REQUIRED
VACUUM LEVEL WITHIN ALLOTTED TIME
CAUSES AND CORRECTIONS:
1. WATER PRESSURE LESS THAN 30 PSIG
-> CHECK WATER SUPPLY PIPING
2. CHAMBER DRAIN STRAINER PLUGGED
-> CLEAN
3. CHECK VALVE MALFUNCTION
-> REPAIR
4. SOLENOID VALVE MALFUNCTION
-> REPAIR
5. DOOR SEAL NOT ACTIVATED
-> CHECK SEAL
-> CHECK SEAL STEAM AND EXHAUST
MORE
HELP
STATUS
PAPER
FEED
EXIT
Содержание AMSCO 400 Series
Страница 2: ......
Страница 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Страница 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Страница 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Страница 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 80: ...6 20 P129394 020 Operator Manual Sterilizer Operation Figure 6 10 Typical Printout of a Liquid Cycle ...
Страница 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...