9-23
Troubleshooting
Operator Manual
P129394-020
Alarm
Description
Screen with Operator Instructions
9.5.2 Chamber
Temperature Probe
Failure
Occurs if chamber
temperature reading is
outside the normal
range.
Screen with Service Instructions
Alarm
Description
Screen with Operator Instructions
9.5.3 Jacket
Temperature Probe
Failure
Occurs if jacket
temperature reading is
outside normal range:
32 - 293°F (0 - 145°C)
STATUS . . ALARM!
211
CHAMBER TEMPERATURE PROBE FAILURE
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
212
CHAMBER TEMPERATURE PROBE FAILURE
-> RTD PROBE, CT, OUTPUT IS OUTSIDE
NORMAL RANGE
CAUSES AND CORRECTIONS:
1. LOOSE CONNECTION IN PROBE WIRING
-> REPAIR
2. PROBE FAILED
-> REPAIR
-> RECALIBRATE
3. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4. MAIN CONTROL FAILURE
-> CHECK CONTROL BOARD POWER STATUS LEDs
-> REPLACE CONTROL BOARD
-> RECALIBRATE
PAPER
FEED
EXIT
STATUS . . ALARM!
213
JACKET TEMPERATURE PROBE FAILURE
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
Содержание AMSCO 400 Series
Страница 2: ......
Страница 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Страница 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Страница 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Страница 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 80: ...6 20 P129394 020 Operator Manual Sterilizer Operation Figure 6 10 Typical Printout of a Liquid Cycle ...
Страница 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...