Ranger S Installation and troubleshooting
31
1 Turn the dial to
ADVANCED
, use the
item selectors
to scroll to item
number
3. FloGuard,
and push the
ENTER
button.
Now you'll see this display:
2
Select
4. Flow Sensor Setup,
and push the
ENTER
button.
Now you'll see this display:
3
Use the
item selectors
to locate the
Custom
option and select it by
pressing the
ENTER
button. Now you'll see the calibration screen:
Note:
A bit of background on how calibration affects the
calculated flow: The Ranger 4000 S needs to know the "K" and
"Offset" values of your sensor (K‐factor and offset are the
parameters of the flow meter known from the datasheet), as
the actual flow will be calculated from this formula: