E-2
MiniCapt® Mobile Operations Manual
Appendix E
: Cleaning Procedures
Runtime Cleaning Guidelines
Runtime Cleaning Guidelines
1.
Remove the impactor head from the sampling location.
2.
Ensure no residuals (e.g. from agar) are present on the impactor head.
3.
Wash the entire stainless steel surface, including mechanical parts, using an
appropriate detergent solution.
•
For example, Veltek Vai’s Dec-Clean, appropriately diluted, typically 0.7
Vol%
(1 fl oz/gallon) in WFI water.
4.
Ensure the impactor head is well cleaned and any cleaning agent residues have
evaporated before proceeding with the autoclave sterilization process.
NOTE:
These procedures are intended to provide general guidelines and may not be
applicable in some specific manufacturing scenarios.
For more information, contact your Particle Measuring Systems local
representative or go to
www.pmeasuring.com
.