MiniCapt® Mobile Operations Manual
20-1
Chapter 20
Troubleshooting
Use this chapter to help troubleshoot issues with the MiniCapt Mobile Microbial Air
Sampler, such as power issues, communication messages, errors and leaks.
Troubleshooting Matrix
Problem
Symptoms
Possible Cause
Try
No Power
No lights
Or
No communications
Power supply not
connected to a power
source
Check the seating of the power
supply cable in the unit.
Verify that the power supply is
plugged in.
Electrical power
source fault
Check the electrical outlet for
power, including any power strips
or surge protectors.
Power supply faulty
Try a different power supply.
Communications not
established – FacilityPro
displays
“Communications Error”
Unable to view data
from working sampler
Batteries are not
charged
Verify batteries are charged.
Ethernet cable
problem
Replace the Ethernet cable that
connects the sampler to an
Ethernet hub.
Unit not configured
correctly
Re-configure the unit. First check
the sensor’s IP Address and the
computer’s IP Address.
Computer hardware or
software problem
Substitute a computer with known
good Ethernet capability.
Refer to your computer’s Ethernet
cared documentation to correct
the fault.