52
--Data<range>:
Data Format: Integer
Data Range: 0~6 (0 is AUTO, 1 is 5mA, 2 is 1mA, 3 is 100uA,
4 is 10uA, 5 is 1Ua, 6 is 0.1uA,)
--Example:
Set the range for IR test in STEP 1 to: 5mA
Command message: FUNC
:
SOUR
:
STEP 1
:
IR
:
RANG 1
--Return Message:
Query message: FUNC
:
SOUR
:
STEP 1
:
IR
:
RANG?
Return to test range of IR in STEP 1, such as 1
Commands for OS Setup Function
FUNC
:
SOURce
:
STEP <sn>
:
OS
:
OPEN
To set /inquiry about the OPEN rate for OS test.
--
Syntax:
Command message: FUNC
:
SOUR
:
STEP <sn>
:
OS
:
OPEN<rate>
Query message:
FUNC
:
SOUR
:
STEP <sn>
:
OS
:
OPEN?
--
Data<rate>:
Data format: integer
Data range: 10~100
Data accuracy: 1
--
Example:
Set the OPEN rate for OS test in STEP1 as 50%.
Command message: FUNC:SOUR:STEP 1:OS:OPEN 50
--Return Message:
Query message: FUNC:SOUR:STEP 1:OS:OPEN?
Return value: 50
FUNC
:
SOURce
:
STEP <sn>
:
OS
:
SHOT
To set /inquiry about the SHOT rate for OS test.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:OS:SHOT<rate>
Query message:
FUNC:SOUR:STEP <sn>:OS:SHOT?
--
Data< rate>:
Data format: integer
Data range: 0~500
(0 is OFF)
Data accuracy: 10
--
Example:
Set the SHOT rate for OS test in STEP1 as 100%.
Command message: FUNC
:
SOUR
:
STEP 1
:
OS
:
SHOT 100
--Return Message:
Query message: FUNC
:
SOUR
:
STEP 1
:
OS
:
SHOT?
Return value:
100
FUNC:SOURce:STEP:OS:STAND
To set /inquiry about the STANDARD capacitance for OS
test.