49
Query message: FUNC:SOUR:STEP 1:AC:TTIM?
Return to the current test time of ACW in STEP 1, such as 1
FUNC
:
SOURce
:
STEP <sn>
:
AC
:
RTIM
To set /inquiry about the RISE time for ACW test.
(
Syntax refer to
AC
:
TTIM)
FUNC
:
SOURce
:
STEP <sn>
:
AC
:
FTIM
To set /inquiry about the FALL time for ACW test.
(
Syntax refer to
AC
:
TTIM)
FUNC
:
SOURce
:
STEP <sn>
:
AC
:
ARC
To set /inquiry about the ARC upper current for ACW
test.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:AC:ARC<current>
Query message:
FUNC:SOUR:STEP <sn>:AC:ARC?
--
Data<current>:
Data format: float
Data range: 0~20.0 mA
(0 is OFF)
Data accuracy: 0.1mA
Data unit: mA
--
Example:
Set the ARC upper current for ACW test in STEP1 as 1mA.
Command message: FUNC:SOUR:STEP 1:AC:ARC 1
--Return Message:
Query message: FUNC
:
SOUR
:
STEP 1
:
AC
:
ARC?
Return to the ARC current upper limit of ACW in STEP 1, such as 1.000
FUNC
:
SOURce
:
STEP <sn>
:
AC
:
FREQ
To set /inquiry about the test frequency for ACW test.
--
Syntax:
Command message:
FUNC
:
SOUR
:
STEP <sn>
:
AC
:
FREQ
<frequency>
Query message:
FUNC
:
SOUR
:
STEP <sn>
:
AC
:
FREQ?
--
Data<frequency>:
Data format: character
Data range: 50/60
Data unit: Hz
--
Example:
Set the test frequency for ACW test in STEP1 as 50Hz.
Command message: FUNC:SOUR:STEP 1:AC:FREQ:50
--Return Message:
Query message: FUNC:SOUR:STEP 1:AC:FREQ?
Return to the test frequency of ACW in STEP 1, such as 50.
FUNC
:
SOURce
:
STEP <sn>
:
AC
:
UNIT1
To set /inquiry about the multiple channels for ACW
test.