26
result processing, test process control, and instrument communication scheme.
Press the “SYSTEM” button to enter, press the “SYSTEM” button again to switch to the sub-
interface.
Note:
After modifying the data in the SETUP interface and SYSTEM interface, you must switch
the interface to save all data modifications.
1.
SYSTEM related sub-interface
Figure 4.2.3.1 Test SYSTEM related interface
Interface parameter description:
Setting Mark
Parameter
value
Description
PASS HOLD
OFF
OFF, PASS signal output time
0.2
~
99.9S
When the test is passed, the hold time of the pass judgment.
STEP HOLD
OFF
OFF, waiting time between steps
KEY
Pause, press ‘START’ key to start the next step
0.1S
~
99.9S
Time to wait between steps during multi-step testing
STRT DLY
OFF
OFF, test wait time
0.1~99.9S
Start and test the waiting time between the first item
GFI
ON, OFF
Ground current detection, enabled.
OFFSET
ON, OFF
Test data clear is enabled.
GET
Get the existing status data as zero data.
FAIL MODE
STOP
If it fails, exit the test state directly.
CONT
If it fails, continue the test and the results can be queried.
PAUS
When the unit fails, the current unit test is suspended
LOOP MODE
ON, OFF
Perform an uninterrupted loop test on existing files.
CTRL MODE
FILE
The file ends and the interface output the test results.
STEP
At the end of each step, the interface outputs a single step
result.