Voltage waiting time
0.3s – 999.9s (Only DC withstand voltage, and meet (rise
time < wait time < (rise time + test time) )
Test time setting
0.3s – 999.9s (at TIMER ON)
Time Accuracy
± (0.2% set value+ 0.1s)
Measuring Function
Keyboard lock
Prevent accidental modification of test conditions or prohibit
modification of test conditions
Base Clearing Function
Current flowing through insulation resistance and distributed
capacitance between output lines can be cleared
Start Wait Delay
At the beginning of the test (press
START
), output the high-
voltage start signal, wait for a certain time, then start the
high-voltage output.
Judgment of Current Overrange
The hardware quickly judges the insulation collapse, which is
faster and safer than voltage sampling, reducing the impact
damage to the product.
ARC Detection
Abrupt signal of sampling current, to determine the potential
hidden dangers of the loop and the approximate scale
Ground current detection
Protect personal safety in case of accidental electric shock or
leakage of high voltage to the outer shell
Alarm volume adjustment
OFF, High, Low
High Voltage Indication
Window Indicator and LED Light Indicator
Storage and Interface
File Programming and Storage
Programmable 20 test files, 16 test steps can be set for AC, DC
and insulation resistance in each file.
USB Interface
Yes
Boot parameter saving
The setting parameters are saved as default parameters, which
can be automatically restored after the next boot.
Control interface
HANDLER, SINGAL
Communication interface
RS232C, USB
2.
Model function correspondence table
ACW
DCW
IR
OS
Test Unit
Scanning Module
S.CK
D.CK
ST9010
*
*
*
*
8channels
ST90101
*
*
ST9010A
*
*
*
*
4channels
ST90101A
*
*
Each scanning module provides four scanning channels for each unit and up to four scanning
modules can be connected.
S.CK and D.CK are dedicated functions of the scanning module.