31
Figure 4.3.1 AC setting interface
AC withstanding voltage (AC) test parameter description:
VOLT
:
0.050
~
5.000kV
AC high voltage test voltage value
UPPER
:
0.001
~
20.00mA
ST9320 series AC withstand voltage current upper limit
0.001
~
10.00mA
ST9010 series AC withstand voltage current upper limit
LOWR
:
0.001
~
20.00mA
ST9320 series AC lower limit current value, must be less than the
UPPER value
0.001
~
10.00mA
ST9010 series AC lower limit current value, must be less than the
UPPER value
OFF
Lower limit is not required
ARC
:
0.1
~
20.0 mA
AC arc current maximum allowable
OFF
No arc is not required
TIME
:
0.2~999.9S
AC withstand voltage test time, the test is end when the time is up
OFF
Test time is not required
RISE
:
0.1
~
999.9S
AC high voltage test voltage rise time
OFF
Default=0.1S
,
Integration with test time
FALL
:
0.1
~
999.9S
AC high voltage test voltage fall time
OFF
Directly cut off voltage output at the end of the test.
(The DUT may
be charged)
FREQ
:
50/60
AC working frequency
4.3.2
DC withstanding voltage test parameter setting
Move the cursor to the test mode, press "F1~F6" to modify the test mode of the current step.
The setting interface is as follows: