![Soredex DIGORA Optime DXR-50 001 Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/soredex/digora-optime-dxr-50-001/digora-optime-dxr-50-001_user-manual_1332635054.webp)
B - 6
Imaging plate unit
B. Installation and setup
Obtain an IP address for the unit from your
network administrator and key it into the IP field in
the
Scanner IP address
area.
NOTE!
The PC and the unit must be in the same
subnet when setting the IP address of the unit.
3.
PC + Unit:
Press and hold down the
Start
key on
the unit and then click the
Send to Scanner
button on the settings window.
You will hear a beep which indicates that the PC
is now sending the IP address the unit. The
Startup
animation will appear on the unit display
4.
PC:
When the
Startup
animation clears click
OK
to connect the PC to the unit.
5. Now connect the other PCs in the network to the
unit. Just enter the IP address into the IP field and
then click
OK
to connect the PC to the unit (it is
not necessary to hold down the
Start
key and
click the
Send to Scanner
button with the other
PCs once the unit has already got ans IP
address).