User’s manual 204253
15
2. Intraoral imaging plate unit
Imaging plate unit
uj7uj
If the exposure time is too short images will be
noisy. Such images may still usable for
somediagnostic purposes.
If the exposure time is too long images will be too
dark or will show patient movement. These
images will not be good enough for accurate
diagnostic examination.
The exposure values table below should only be
used as a guide.
Recommended exposure values (s) for DC x-ray units*
60 kV, 7 mA
70 kV, 7 mA
9” cone
12” cone
9” cone
12” cone
Adult
Child
Adult
Child
Adult
Child
Adult
Child
Bitewing
0.250 0.160 0.500 0.300 0.125 0.080 0.250 0.160
Maxillary incisor
0.200 0.125 0.400 0.250 0.100 0.063 0.200 0.125
Maxillary cuspid
0.200 0.125 0.400 0.250 0.100 0.063 0.200 0.125
Maxillary molar
0.300 0.200 0.600 0.400 0.160 0.100 0.300 0.200
Occlusal
0.250 0.160 0.500 0.300 0.125 0.080 0.250 0.160
Mandibular incisor 0.200 0.125 0.400 0.250 0.100 0.063 0.200 0.125
Mandibular cuspid 0.250 0.160 0.500 0.300 0.125 0.080 0.250 0.160
Mandibular molar 0.250 0.160 0.500 0.300 0.125 0.080 0.250 0.160
*For AC x-ray units increase the exposure times by about 30%
3.
Protect yourself from radiation
and take the
exposure.
4. After the exposure remove the sealed hygiene
bag from the patient’s mouth.
CAUTION
:
If there is a risk of cross contamination, wash,
disinfect and dry the hygiene bag before opening
it.