![Soredex DIGORA Optime DXR-50 001 Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/soredex/digora-optime-dxr-50-001/digora-optime-dxr-50-001_user-manual_1332635031.webp)
User’s manual 204253
25
3. Setup options
Imaging plate unit
3.2 Resolution
1. In the
Image Scanning
section, select either
Super
or
High
(default).
Super
gives a pixel size of 35
μ
. This results in
images with better resolution, but more memory is
required as the image files are larger.
High
a pixel size of 64
μ
. This results in images
with less noise especially if short exposure times
are used.
2. Click
OK
to close the window and accept the
selected resolution.
3.3 Image Processing
1. In the
Image processing
click the
Noise Filter-
ing
check box.
Noise filtering is normally always selected. It
makes images smoother when they are taken at
lower radiation doses.
2. Click
OK
to close the window and select the noise
filtering.
Содержание DIGORA Optime DXR-50 001
Страница 2: ......
Страница 40: ...34 User s manual 204253 Imaging plate unit 7 Warnings and precations...
Страница 43: ...User s manual 204253 A 3 A Technical Specifications Imaging plate unit A 3 Main dimensions...