User’s manual 204253
17
2. Intraoral imaging plate unit
Imaging plate unit
As soon as the protective cover is removed the
unit detects that the IP is in the unit door and will
switch from the standby state to the ready state
(ready to use) and automatically slide the IP into
the unit.
CAUTION:
If the IP does not slide into the unit after the
protective cover has been removed, the IP is
misaligned or has been placed in the unit the
wrong way round. Reposition the IP and insert it
into the unit again.
If after repositing the IP it still does not slide into
the unit press the
Start
key to manually switch the
unit from the standby to the ready state and the IP
will then slide into the unit when the protective
cover is removed.
The busy animation will appear on the display
which indicates that the imaging plate is being
read.
After few seconds a preview image will appear on
the unit display
PC:
A read-out progress window will appear on
the PC display. After a few seconds the image will
appear in the dental imaging software.
The image can now be saved. Refer to the
documentation supplied with the dental imaging
software you are using.