B. Installation and setup
Imaging plate unit
Appendix B. Installation and setup
Installation and setup must only be carried out
by service personnel trained and approved by
the manufacturer of the unit.
1. System installation
Positioning the unit
Position the unit on a stable flat surface so that
vibrations will not degrade the image quality.
The unit can also be attached to a wall, under or on a
shelf using the optional
scanner mounting kit
(pt.
no. 204138).
Do not position the unit in direct sunlight or near bright
light. Sunlight or bright light must not be allowed to
shine directly on the unit door into which the IPs are
inserted.
The unit must not be positioned so that it touching
other equipment. The unit must not be placed on top
of or under other equipment.
The unit can be positioned within the environment in
which the patient is examined and treated (patient
environment).
Positioning the PC(s)
(not supplied)
The PC(s) connected to the unit should not be used in
the patient environment.
The minimum horizontal distance between the patient
and the PC(s) is 1.5 m (4.5 ft).
The minimum vertical distance between the patient
and the PC(s) is 2.5 m (6.5 ft).