Imaging plate unit
2. Intraoral imaging plate unit
3.
PC:
To release the unit click the
GREEN
multiconnect icon and then click the
Free the
scanner
(unit) button.
Removing IPs from the plate collector
When removing IPs from the plate collector hold them
by their edges. Alternatively, pull the plate collector out
of the unit and tip the IPs out onto a flat clean surface.
If they are not to be reused immediately or within a
short period of time, store them in their storage box.
Standby mode
If no IP is inserted into the unit within a certain period
of time, the unit will beep several times and the status
light will start to flash. When it turns yellow the unit is in
the standby mode.
The unit will automatically exit the standby mode as
soon as a new imaging plate is inserted into the unit
or when the
Start
key is pressed.
After four hours the unit will exit the standby mode and
automatically shut down.