User’s manual 204253
11
2. Intraoral imaging plate unit
Imaging plate unit
When the status light turns green and ready
animation, indicating IP insertion, appears on the
unit display, the unit is ready to use (in the ready
state).
NOTE:
If the ready animation does not appear, check the
system setup described in the installation
instructions.
Preparing an IP for exposure
IMPORTANT NOTE
If the IP is being used for the very first time or if it has
not been used within the the last 24 hours, it must be
erased before use to remove any fogging caused by
background radiation. See section,
Erasing an
imaging plate
.
CAUTION:
When handling IPs, protective covers and hygiene
bags take all appropriate measures and precautions
to prevent cross contamination.
1. Place the IP you wish to use onto a protective
cover. The
light blue
side (sensitive) of the IP
must face and be placed on the half of the
protective cover that is the same shape as the IP.
2. Fold the half of the protective cover with the
semicircular cutout over onto the IP. The metal
disk on the back (black side) of the IP must
appear in the semicircular cut out.