18
User’s manual 204253
Imaging plate unit
2. Intraoral imaging plate unit
CAUTION:
If the metal disk on the rear of the IP can be seen
on the digital image, it indicates that the IP was
exposed from the wrong side.
CAUTION - RETRIEVING IMAGES
If the image is not transferred to the PC because
of a network, PC or software failure, the image
can be retrieved from the unit’s memory as long
as the unit is
NOT
switched off.
For information on how to do this see section
Retrieve last image
.
5. After the IP has been read it will be automatically
erased and then ejected from the unit into the
plate collector.
After the IP has been ejected the ready animation
will reappear on the display. The unit is now ready
to read the next IP.