Imaging plate unit
2. Intraoral imaging plate unit
Reading an imaging plate
Single user configuration
NOTE:
If the unit is operated in a multiconnect configuration,
please refer to
Multiconnect configuration
first to
reserve the scanner.
1. Pull the flap to open the hygiene bag.
2. Keep the IP in the protective cover so that you do
not touch the IP or allow it to be exposed to
ambient light, and then slide the protective cover
and IP out of the hygiene bag.
3. Hold the protective cover and IP so that the white
side of the protective cover is on the right. The
metal disk on the IP will also be on the right.
Insert the protective cover and IP into the unit
door.
A magnet will hold the IP in the correct position.
4. Slide the protective cover off of the IP, and leave
the IP in position half way in the unit door.