7. Warnings and precations
Imaging plate unit
•
Unit not suitable for use in the presence of flammable anaesthetic mixture with air
or with oxygen or nitrous oxide.
•
For ethernet connections, use an unshielded CAT6 LAN cable, so that multiple
chassis must not be connected! The PC / Ethernet switch to which unit is
connected to, should be approved appropriately (e.g. EN 60950, IEC 60950, UL
60950). After installation check that the IEC 60601-1 leakage current levels are
not exceeded.
•
In order to maintain safe and correct functioning of the unit, only the power supply
unit (PSU) delivered with the unit or distributed by authorized dealers. Please
refer to the unit technical specifications for a list of the PSUs.
•
If this device will be used with 3rd party imaging application software not supplied
by the manufacturer, the 3rd party imaging application software must comply with
all local laws on patient information software. This includes, for example, the
Medical Device Directive 93/42/EEC and/or FDA if applicable.
•
Medical electrical equipment needs special precautions regarding EMC and
needs to be installed according to EMC information.