22
User’s manual 204253
Imaging plate unit
2. Intraoral imaging plate unit
Shutting down the unit
1. Press and hold power on / off key until the
indicator light goes off.
NOTE
:
If there is an untransferred image in the unit’s
memory the unit cannot be shut down. The image
must be transferred first. Refer to section
Retrieve last image
.
IP erasing mode (Initial erasing of the IPs)
If an IP is being used for the very first time or has not
been used for 24 hours and has not been exposed to
ambient light the erasing procedure must be carried
out before the IP is used. The erasing procedure
removes any fogging due to background radiation.
The dedicated erasing mode erases the IP but does
not send the image (blank) to the dental imaging
software.
1. Press and hold down the
Start
key for several
seconds until the erase mode animation appears
on the unit display. The unit is now in the erase
mode.
2. Hold the IP by its edges (or use a protective
cover) and position it so that the side with the
metal disk (the back) is on the right.
Insert the IP into the unit door (remove the
protective cover if used). A magnet will hold the IP
in position and then automatically slide it into the
unit.