1 Introduction
1-3
PVA-600 Specifications
Electrical Specifications
Safety Rating: Measuring Category CATIII 600V.
Table 1. PVA-600 electrical specifications
Parameter
Specification
Current Measurement Range
1
0 to 20 A dc
Voltage Measurement Range
0 to 600 V dc
Minimum Voc
20 V dc
Minimum Isc
1 A dc
Wireless Communications Range
10 m (industrial building walls) to 75 m (open
range)
Measurement Sweep Time
2
80 ms to 240 ms
Measurement Points per Trace (typical)
100
PV Models
Sandia (>420 modules)
5-Parameter (>1760 modules)
Simple Datasheet Model (user enters datasheet
values)
Battery Life
≈20 hours (normal use)
Charging Time
6 hours
Operating Temperature
+0º C to +50º C
Storage Temperature
-20º C to +60º C
Operating Humidity
The normal humidity range is 80% relative
humidity for temperatures up to 31ºC,
decreasing linearly to 50% at 40ºC. Higher
humidity levels should not affect the
performance or safety of the PVA-600.
1
Conventional PV modules and strings may be measured in parallel, up to the current
limit specified here. High-efficiency modules should NOT be measured in parallel.
2
Automatically selected. Measurement sweep time depends upon the characteristics of
the test device (PV module, string, or array).