1 Introduction
1-5
PVA Sensor Kit Specifications (Optional)
Table 3. PVA Sensor Kit temperature specifications
Parameter
Specification
Thermocouple Type
K
Range
-100º C to 1260º C (-148º F to 2300º F)
Resolution
1º C
Accuracy (greater of)
±0.5% of rdg or ±1.0º C (1.8 º F)
Table 4. PVA Sensor Kit irradiance specifications
Parameter
Specification
Sensor Type
Silicon Solar Cell
Angle of Incidence Effect
Typically < 1% if AOI is less than 60 degrees from
normal
Temperature Effect
Typically < ±2.7% variation over -20º C to +70º C
Uncertainty Contributed by Wireless Link
±0.5% or reading
Uncertainty from Time Lead or Lag in
Reading Irradiance Sensor
0.5% to > 5% depending on atmospheric conditions
Содержание PVA-600
Страница 1: ...Solmetric PVA 600 PV Analyzer User s Guide ...
Страница 10: ...x The page left blank intentionally ...
Страница 18: ...1 Introduction 1 8 The page left blank intentionally ...