4 Using the PVA-600 Software
4-9
Name
Description
No USB Wireless
(indicator)
Displayed when the USB Wireless Adapter is not found. When
displayed, clicking on this indicator accesses information to aid in
troubleshooting the problem.
No I-V Unit (indicator)
Displayed when communication between the I-V Measurement
Unit and PC is not established. When displayed, clicking on this
indicator accesses information to aid in troubleshooting the
problem.
NOTE
– The most common reasons for this state include
out-of-range, or the I-V Measurement Unit is turned
off.
Ready (indicator)
Displayed when the system is ready to start a new I-V
measurement.
Measure Now
button
Highlighted when the system is ready to start a new I-V
measurement. Each time this button is clicked, a single I-V
measurement sweep is started.
Measuring (indicator)
During a measurement, the
Measuring
label blinks.
Paused (indicator)
Displayed when the I-V measurement process is paused by pressing
the button on the I-V Measurement Unit. In this state, PV source
connections may be changed without interrupting a measurement.
WARNING - PV circuits continue to present danger of
electrical shock while system is paused.
FAILURE TO TAKE APPROPRIATE
SAFETY PRECAUTIONS COULD LEAD
TO PERSONAL INJURY OR LOSS OF
LIFE.
Disabled (indicator)
Displayed when a problem exists other than communication
between the PC and I-V Measurement Unit. Problems could be
related to low battery, over current, over voltage, over temperature,
reversed polarity, etc. No measurements can be taken while in this
state.
When displayed, clicking on this button accesses information to aid
in troubleshooting the problem.
?
button
Clicking on this button accesses information specific to the current
state of the PVA-600.
Snapshot
button
Takes the current measured values of the data and populates one of
the columns in the right hand table.
Clear Last
button
Clears just the most recent section of data from the Snap Shots
table.
Clear All
button
Clears all columns of data from the Snapshots table.