Clinician Guide
p. 40
TROUBLESHOOTING
Issue
Possible Solution
A continuous two-tone alarm is
sounding, and the amber light is
lit or flashing.
Delivery has stopped. Read the message on the display and
refer to the list of messages in the table below. If the display is
blank or contains random characters, the 4 AA batteries or the
rechargeable battery pack may be depleted. Install 4 new AA
batteries or a rechargeable battery pack.
The pump is sounding 2 beeps
every 2 seconds, and the amber
light is flashing.
Look at the message on the display and refer to the list of
messages in the table below.
Three beeps sound every 5 minutes.
This is a reminder that the pump is stopped.
After installing 4 new AA batteries
and powering up the pump, no
screen appears and no beep sounds.
The batteries may be installed incorrectly. Review the
procedure for installing batteries. Be sure to match the polarity
(+ and –) markings inside the battery door with the markings
on the batteries. If there is still no power, the batteries may be
completely depleted.
ALARMS AND MESSAGES, ALPHABETICAL LIST
A more detailed list of alarms and messages is included in the
Operator’s Manual.
Alarm / Message
Alarm
Priority
Description / Corrective Action
(Screen is blank and
alarm is sounding)
High
The pump was delivering and the batteries were removed or the
battery door was opened. The pump lost power and is no longer
delivering. Clear the alarm by turning the pump back on, or the
alarm will stop after the power has been off for a minimum
of 2 minutes.
A setting was edited,
but not saved, and the
edit was lost.
Medium
A parameter was being manually edited, but it was not
saved and the pump reverted to the home screen. Select
Acknowledge to clear the alarm and if appropriate, edit the
parameter and save.
Air in-line detected.
Press Acknowledge
then prime tubing.
High
The air detector has detected air in the fluid path. The fluid
path may contain air bubbles. The pump was delivering and
is now stopped and will not run. Select Acknowledge to clear
the alarm.
If the fluid path contains air bubbles, close the clamps
and disconnect the fluid path from the patient. Follow the
instructions in the Operator’s Manual for removing air by
priming. Restart the pump.
Battery depleted.
Pump stopped.
High
The pump was delivering but is now stopped and the battery
power is too low to operate the pump. If the AC adapter is
attached, select Acknowledge to clear the alarm.
Remove the batteries. Install 4 new AA batteries or a
rechargeable battery pack. To start delivery, good batteries
must always be installed even when an external source of
power is connected. If appropriate, restart the pump.