Clinician Guide
p. 8
SECURITY SETTINGS
Security settings are used to limit
patient and unauthorized access to
certain programming and operating
functions of the pump. Pump functions
are protected by three different security
codes that can be customized.
Keypad Code - The keypad code is
for clinicians who need to modify and
review patient-specific parameters
and manage air and occlusion
parameter settings.
Clinician Code - The clinician code
allows access to all the functions the
keypad code allows. In addition, it
allows users to modify most advanced
task settings, format the time and
date, change delayed start and
priming security options, and select
new protocols.
Administrator Code - The administrator
code allows access to all pump
functions and allows users to change
protocol ranges for delivery settings,
reset the pump to factory defaults,