Clinician Guide
p. 28
Removing a Cassette
1.
Make sure the pump is stopped
before removing the cassette
2.
Close the tubing clamp
3.
If locked, insert the pump key and
turn the cassette/keypad lock
counter-clockwise into the unlocked
position. “Cassette Unlocked”
briefly appears in the status bar
Note: Only PCA infusions require the
cassette to be locked in order to run
the pump.
4.
Push down on the cassette latch
until cassette detaches
Attaching a Cassette
Obtain a new, filled CADD™ medication
cassette reservoir or CADD
®
administration set attached to
a flexible IV bag.
Before attaching a new cassette, make
sure the pump is powered on. Once
the cassette is attached, the pump
automatically displays screens that
allow you to verify the cassette type,
prime the fluid path, reset the reservoir
volume, review pump settings, and/or
start the pump.
1.
Clamp the tubing
2.
Make sure the cassette latch is
unlocked then open the latch
3.
Insert the cassette hooks into the
hinge pins on the bottom of the
pump, then swing the cassette to
the latch position
4.
Place the pump upright on a firm,
flat surface, and then press down
on the latch side of the pump so
the cassette fits tightly against
the pump