Clinician Guide
p. 12
START NEW PATIENT -
SINGLE PROTOCOL
To start a new patient using the
single protocol that was downloaded
to the pump:
1.
Power on the pump. The pump will
display the home screen
2.
Attach a cassette. (see page 25
for instructions)
3.
You will be prompted to prime
the tubing. (see page 26
for instructions)
4.
Depending on your facility’s
policies and procedures, one
of two messages will appear
on the screen:
a.
“Review pump settings”
may appear on the screen
•
Press the right soft key Review
•
Unlock the keypad, if required,
the delivery settings will appear
•
Review and accept each delivery
setting by pressing the right soft
key Accept Value
•
If any of the delivery setting
values need to be edited, press
the Select key
•
Once all the values have been
accepted, press the right soft
key Next
•
“Start Pump?” appears,
choose Yes
b.
“Start Pump?” may appear on
the screen
Choose No. The home screen will
appear. To review the delivery
settings before starting the
infusion:
•
Press the right soft key Tasks
•
Scroll to highlight View Delivery
Settings, press Select
•
Review each setting by scrolling
through each parameter
•
Press Select to edit any
specific parameter. Unlock
keypad if required
•
When all parameters have been
reviewed, press the left soft
•
key Back until you return to the
home screen
•
Now press the Stop/Start key,
“Start Pump?” appears
•
Choose Yes
Programming the Pump