CADD
®
-Solis VIP Pump
p. 37
The following functions are in the Advanced Tasks menu:
Advanced Task Menu Function
Description on Page
Give Clinician Bolus (PCA Therapy)
17
Step Up (Step therapy)
23
Step Down (Step therapy)
23
Patient Permissions
37
Air and Occlusion Settings
See Operator’s Manual
Alarm Settings
See Operator’s Manual
Security Settings
8
Start New Patient
12
Start New Protocol Same Patient
13
Delivery Hard and Soft Limits
See Operator’s Manual
Reset to Factory Defaults
See Operator’s Manual
To access any item in the Advanced
Tasks menu:
1.
From the home screen, select Tasks
2.
Press
▲
or
▼
to highlight View
Advanced Tasks and press Select
3.
Press
▲
or
▼
to highlight the
Advanced Tasks you wish to view/
edit and press Select
4.
Unlock the keypad, if required
5.
View/edit the setting and press
either Save to confirm the edit or
Don’t Save to exit the task
PATIENT PERMISSIONS
Security levels are used to limit
unauthorized access to certain
programming and operating functions.
In some cases, the default security
settings may be changed to allow
patients additional access to pump
functions. Do not reduce security
settings without providing proper
training to patients.
The Patient Permissions task allows
you to control a patient’s access to two
specific tasks: priming and setting a
delayed start. Setting the permission to
“Off” allows the patient to access one
or both tasks without having to enter a
security code. Setting the permission
to “On” restricts the patient’s access to
one or both tasks because a security
code is required.
There are two items under
patient permissions:
Priming Security On/Off - Setting
this value to “Off” allows the patient/
caregiver to prime the tubing on the
pump without having to enter a security
code. If this value is set to “On”, the
pump will ask for a security code before
allowing the user to use the priming
function on the pump.
Delayed Start Security On/Off - Setting
this value to “Off” allows the patient/
caregiver to set Delayed Starts or Next
Dose Start Times without having to
enter a security code. If this value is set
to “On”, the pump will ask for a security
code before allowing the user to enter
and complete this task.