START NEW PATIENT –
PROTOCOL LIBRARY
To Start a New Patient Using a Protocol
From a Library on the Pump:
1.
Make sure the pump is stopped
2.
In the Advanced Tasks menu,
press
▲
or
▼
to highlight Start
New Patient or Start New Protocol
Same Patient and press Select
3.
The next screen informs you
completing this task will
overwrite all delivery settings.
Press continue to unlock the
keypad and continue programming
the pump
Note: To edit individual settings rather
than starting a new patient or protocol,
see page 11 for more information on
editing delivery settings.
4.
Unlock the keypad
5.
Press
▲
or
▼
to highlight a therapy
and press Select
6.
Press
▲
or
▼
to highlight a qualifier
and press Select
7.
Press
▲
or
▼
to highlight a drug
and press Select
8.
A screen appears with a summary
of your selections. Review them to
confirm they are correct. Select Yes
9.
The “Review pump settings” screen
appears. Select Review
10.
Press
▲
or
▼
to scroll through and
review all the settings. Press Accept
Value if the highlighted setting is
correct or press Select to edit the
value. Then press Save
11.
When the review is complete,
select Next
12.
If a cassette is attached, latched
and locked (if using the PCA
therapy), the “Prime tubing?”
screen may appear. The “Start
pump?” screen appears. Select Yes
to start the pump immediately or
No start the pump at a later time
Manual Mode Programming
You can manually program the
CADD
®
-Solis VIP pump if the
physician’s orders do not match the
single protocol or any of the protocols
in the library or if the patient’s therapy
requires wider programming ranges
than available in a standard protocol.
Unlike protocols that are created
and downloaded into the pump by
the CADD
®
-Solis medication safety
software, manual programming
does not contain any programming
limits. For more information, see the
Operator’s Manual or your CADD
®
-Solis
system administrator.
DELAYED START / NEXT DOSE
START TIME
Delayed Start
Note: This task is available for all
therapies except intermittent.
The Set Delayed Start task allows you
to delay the start of an infusion by up to
96 hours by selecting the date and time
the infusion should begin.
If the Set Delayed Start task is
programmed, the start time is
displayed as the last delivery setting.
This is the time the next infusion will
begin. The pump must be running on
the selected date and time in order for
delivery to begin.
If the Set Delayed Start is programmed,
the KVO rate is active until the
infusion begins.